Return Loss Measurement
A Background to Return Loss Measurement
192
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
Measuring the Power Transmitted Through
the DUT
Remove the termination from the DUT and connect the DUT to a Power
Meter so that the power that is transmitted through the DUT can be
measured. This power is called E
il
Figure 116 Measuring the Power transmitted through the Device Under Test
You can use E
il
to calculate the insertion loss, see “
Calculating the
Insertion Loss of the DUT
” on page 196.
Calculating the Return Loss of the DUT
The system may be represented by the general diagram shown below:
The reflected power, measured by the instrument, (P), from the component
with the known reflection factor, is given by the sum of:
• the part of the power, reflected by the component, which is transmitted
through the coupler, and
• the reflections due to the measurement system.
M
il
measurement
patchcord
DUT
Power Meter
E
il
Agilent 81610A
Reflectivity
R
DUT
Reflection-free
Termination
R
M
ref
P
ref
Agilent 81610A
Reflectivity
P
in
P
r
P
src
t
1
k
1
k
2
t
2
Summary of Contents for 8163B
Page 4: ......
Page 20: ...20 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 22: ...22 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 62: ...Getting Started A Sample Session 62 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 368: ...Performance Tests Test Record 368 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 378: ...Performance Tests Test Record 378 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 465: ......