Error Messages
Tunable Laser Sources
394
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
If one of the selftests fails, please try to reboot the instrument. If the selftest
continues to fail, please return the mainframe to your local service center for
repair.
error message
Initialization Error (TASK_VERIFY, id1, id2)
description
A task could not be started
error value
id1
task name
id2
maximum available memory
action
Reboot the instrument. If the failure persists, please return the unit to
your local service center for repair
error message
Initialization Error (Tcfs Amplifier Calibration failed, error value)
description
The calibration of the tuning control feedback system amplifier has
failed
error value
0
(not significant)
action
Please return the unit to your local service center for repair
error message
Initialization Error (TcfsAdpEEPROMInit, error value)
description
Error reading tuning control feedback system adapter board EEPROM
error value
1
(not significant)
action
Please return the unit to your local service center for repair
error message
Initialization Error (TcfsEEPROMInit, error value)
description
Error reading tuning control feedback system EEPROM
error value
1
(not significant)
action
Please return the unit to your local service center for repair
error message
Initialization Error (Tcfs temperature not settled, error value)
description
Temperature has not settled at the tuning control feedback system
error value
0
(not significant)
action
Please return the unit to your local service center for repair
error message
Initialization Error (TmpCalDataInit, error value)
description
Error accessing caldata
error value
1
(not significant)
action
Please return the unit to your local service center for repair
Summary of Contents for 8163B
Page 4: ......
Page 20: ...20 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 22: ...22 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 62: ...Getting Started A Sample Session 62 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 368: ...Performance Tests Test Record 368 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 378: ...Performance Tests Test Record 378 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 465: ......