Performance Test Instructions
Performance Tests
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
355
Display/Key Functional Test
1
Setup the test equipment as shown in Figure 201 :
Insert the fixed laser source module in slot 1 and the power sensor in slot 2. If
you use the Agilent 8164A/B Lightwave Measurement System, insert the back-
loadable Tunable Laser module, if available, in slot 0.
2
Switch on the mainframe and wait until it has booted.
Testing Hardkeys
1
Press Config. You should see the Configuration menu.
2
Press Aux. You should hear a “beep” and the Configuration menu should
disappear.
3
Press Appl. You should hear a “beep”.
4
Press Help. You should hear a “beep”.
5
Press Preset, the green hardkey. An egg-timer icon should be displayed while
all parameters are set to their default values. The Overview Screen showing all
channels should appear.
6
Press Channel several times.
The highlighted field on the display should toggle between the channels you
have inserted a module into.
Testing the Softkeys and the Cursor Key
1
Press the right arrow of the cursor key several times.
Check the fields on the display being highlighted one after another.
2
Press the left arrow on the cursor key several times.
Check the fields on the display being highlighted one after another.
3
Press the up arrow on the cursor key several times.
Check the fields on the display being highlighted one after another.
4
Press the down arrow on the cursor key several times.
Check the fields on the display being highlighted one after another
5
Using the cursor key, move to the power parameter, [P], of the Power Sensor
module.
Testing the Softkeys and the E
NTER
Key
1
Press [Menu]. The menu should appear.
2
Press [Close]. The menu should disappear.
3
Press [Details]. The details screen should appear. You should see the [Cal],
[Ref], and [Range] parameters in addition to [P] and [
λ
]. If you use the
Summary of Contents for 8163B
Page 4: ......
Page 20: ...20 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 22: ...22 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 62: ...Getting Started A Sample Session 62 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 368: ...Performance Tests Test Record 368 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 378: ...Performance Tests Test Record 378 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 465: ......