How to Measure Power
Power Measurement
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
93
• a source is connected to the Power Meter’s input connector,
• the fiber connected to the Power Meter’s input connector is collecting light, or
• the Power Meter receives ambient light because the input connector is
uncovered.
Press Enter and start again at step 1.
4
When the zeroing operation finishes, press the [Close] softkey to close the
menu.
Figure 46 Zeroing Fails, if the Power Meter Receives Input Light.
If you see the screen shown in Figure 46 , the zeroing operation has failed
because the Power Meter received light. The most common reason for
zeroing to fail is if:
N O T E