Return Loss Measurement
Return Loss Measurement
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
185
Measuring the Insertion Loss
To measure insertion loss, you measure the power transmitted through the
DUT, as shown in Figure 107 .
As you have already, in Step 2 on page 182, set the [Ref] parameter to the
power transmitted through the Reference Cable. The displayed power in
dB is equal to the insertion loss.
Set the Front Panel Delta as described on “
Front Panel Delta Calibration
”
on page 181.
N O T E