Error Messages
Screen Status Messages
382
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
Screen Status Messages
Message
Meaning
Description
8163A/B
8164A/B
SL
SLock
Security Lock
The module is locked for security reasons. No measurement is possible
until you unlock the module
RiL
RiLock
Remote InterLock
The remote interlock connector of your mainframe is open. No mea-
surement is possible until the connection at the remote interlock con-
nector closes
NoH
No Head
No Head
There is an optical head interface module inserted in the mainframe
that does not have an Optical Head connected to it. No measurement
is possible using this channel until you:
- reconnect an optical head or
- deselect this optical channel as a power measurement channel
NoD
No Data
No Data
No data has been measured or measurement is not finished. You can-
not access the analysis screen yet. Printout and file operations are not
yet possible.
Set
Settling
Settling
A tunable laser source module is settling. The instrument is busy and
no measurement can be performed yet.
NoR
No Ref
No Reference
No reference sweep has been performed. Loss will be calculated using
nominal laser power as a reference.
B
Busy
Busy
Module is busy performing last operation.
Summary of Contents for 8163B
Page 4: ......
Page 20: ...20 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 22: ...22 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 62: ...Getting Started A Sample Session 62 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 368: ...Performance Tests Test Record 368 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 378: ...Performance Tests Test Record 378 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 465: ......