Return Loss Measurement
Return Loss Measurement
172
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
Setup
External and Internal Sources
The Return Loss measurement setup described uses:
• An Agilent 81654A Source module, inserted as a second module in the the
same mainframe as the Return Loss module, or
• An internal source (Agilent 81611A/2A/3A/4A Return Loss modules only).
Making sure all the connectors are clean, set up the instrument as shown
in Figure 88 if you are using an External Source,
Figure 88 Return Loss Measurement Setup - External Source used
or Figure 89 if you are using an internal source
(Agilent 81611A/2A/3A/4A Return Loss modules only).
Figure 89 Return Loss Measurement Setup - Internal Source used
• If an external source is used, connect it to the Return Loss module Input.
• Attach the high return loss connector of the patchcord to the Return Loss
module Output. The high return loss connector is the connector with the
orange sleeve.
Agilent 81113PC
8161x
8163B Lightwave Multimeter
Laser
Source
Agilent 81113PC
8161x
8163A/B Lightwave Multimeter
Agilent 81109AC
8161x
8163B Lightwave Multimeter
If you are using a Fabre-Perot source, you must fix its output cable to
ensure minimum cable movement.
N O T E
Summary of Contents for 8163B
Page 4: ......
Page 20: ...20 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 22: ...22 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 62: ...Getting Started A Sample Session 62 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 368: ...Performance Tests Test Record 368 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 378: ...Performance Tests Test Record 378 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 465: ......