53220A/53230A User’s Guide
167
Triggering and Gating
5
fore, the duration of the gate, is set or changed with the command:
[SENSe:]TOTalize:GATE:POLarity {POSitive|NEGative}
[SENSe:]TOTalize:GATE:POLarity?
(query form)
POSitive
starts totalizing on a positive edge on the Gate In/Out BNC or chan-
nel 1/channel 2 input, and stops totalizing on the
next
negative edge.
NEGa-
tive
starts totalizing on a negative edge on the Gate In/Out BNC or channel
1/channel 2 input, and stops totalizing on the
next
positive edge.
CONFigure
and
MEASure
do not change the polarity setting. Following a reset
(
*RST
) or instrument preset (
SYSTem:PRESet
or
Preset
key), the negative
edge is selected.
To set the polarity of the external gate signal:
// select the gate polarity and source
CONF:TOT:TIM
SENS:TOT:GATE:POL POS
// set polarity
SENS:TOT:GATE:SOUR EXT // set source
External Gate Signal Threshold