156
53220A/53230A User’s Guide
5
Triggering and Gating
moves the counter to the
triggered
state and the beginning of the gate cycle.
The counter remains in the
triggered
state until the sample count (readings
per trigger) is reached. The counter then returns to the
wait-for-trigger
state
until the next system trigger is received. The counter returns to the
idle
state
once the product of the trigger count x sample count is reached.
Measurement Gate
Control of the measurement gate allows you to select the duration of the mea-
surement. The gating sequence of the trigger/gate cycle begins after a system
trigger signal is received, and is repeated for
each
measurement specified in
the sample count.
Gate Set Up
The counter’s
gate source
determines the path through the gate cycle and the
related parameters that must be considered. A gate source is required for all
frequency, totalizing, and time interval measurements.
Although selection of the gate source is shown at the beginning of the cycle
N O T E
The gate sequence occurs while the counter is in the
triggered
state (Figure 5-2).
Although shown in the figure after the counter is initiated, gate
configuration
like system trigger
configuration,
occurs while counter is in the
idle
state.
N O T E
Counter measurements are based on configurations comprised of multiple param-
eters. The
easiest
and most common starting point for setting these parameters
programmatically
is using commands within the CONFigure and MEASure sub-
systems (Chapter 3). These subsystem commands are considered “high-level”, as
multiple counter parameters are set or defaulted from a single command. “Low-
level” commands such as the gating commands described here allow you to
change selected gating parameters without altering other areas of the counter
configuration.
Summary of Contents for 53220A
Page 10: ...10 53220A 53230A User s Guide Contents...
Page 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Page 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Page 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Page 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Page 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Page 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...