53220A/53230A User’s Guide
5
Triggering and Gating
The system trigger is configured using the commands within the
TRIGger
and
SAMPle
subsystems. Figure 5-3 show the sequence in which the com-
mands are commonly used.
The system trigger source which starts the trigger/gate cycle is set with the
command:
TRIGger:SOURce
{IMMediate|EXTernal|BUS}
TRIGger:SOURce?
(query form)
- trigger source
IMMediate
sets a continuous trigger signal. By default,
CON-
Figure
sets the trigger source to
IMMEDiate.
- trigger source
EXTernal
sets the trigger source to an external trigger applied
to the rear panel ‘
Trig In
’ BNC connector.
- trigger source
BUS
causes the instrument to be triggered by the
*TRG
com-
mand received over an IO interface.
Trigger Source Example
//configure for frequency, set system trigger parameters
CONF:FREQ 5E6,0.1,(@2)
TRIG:SOUR EXT
// trigger source external
System Trigger Slope
When the system trigger source is set to
EXTernal
, the triggering slope (edge)
of the signal is set with the command:
Summary of Contents for 53220A
Page 10: ...10 53220A 53230A User s Guide Contents...
Page 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Page 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Page 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Page 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Page 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Page 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...