53220A/53230A User’s Guide
101
53220A/53230A Measurements
3
MEASure:PDUTycycle? [{<reference>|MINimum|MAXimum|
DEFault}] [,<channel>]
CONFigure:PDUTycycle [{<reference>|MINimum|MAXimum|
DEFault}] [,<channel>]
MEASure:NDUTycycle? [{<reference>|MINimum|MAXimum|
DEFault}] [,<channel>]
CONFigure:NDUTycycle [{<reference>|MINimum|MAXimum|
DEFault}] [,<channel>]
-
reference
specifies the threshold level where the duty cycle measurement
begins. For positive duty cycle measurements, the level at which the signal
passes through the threshold in the positive (rising) direction and again in the
negative (falling direction) defines the positive duty cycle. Similarly, the level
at which the signal passes through the threshold in the negative (falling) direc-
tion and again in the positive (rising) direction defines the negative duty cycle.
To specify a threshold level
from
10%
to
90% of the
peak-to-peak input volt-
age
, use a numeric value with
no
suffix or with the suffix PCT (e.g. 30 or
30 PCT).
To specify a threshold level in
absolute voltage
within the ranges:
5.125V range: -5.125V to +5.125V
51.25V range: -51.25V to +51.25V
use a numeric value with the V or MV (millivolt) suffix: 100 MV or .1V (spaces
allowed).
If
reference
is omitted or specified in percent, auto-leveling is enabled. If
specified in absolute voltage, auto-leveling is disabled
-
channel
is counter channel 1 or 2 specified as
(@1)
or
(@2)
.
Summary of Contents for 53220A
Page 10: ...10 53220A 53230A User s Guide Contents...
Page 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Page 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Page 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Page 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Page 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Page 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...