53220A/53230A User’s Guide
3
53220A/53230A Measurements
Setting the Measurement Time Out
The measurement time-out is the time allowed for
each
measurement to
complete. If a measurement
does not
complete before the time out expires,
9.91E37
(not a number) is returned and the display indication is: - - - - - - -.
The sequence continues with the next reading in the sample count.
Specifying a time out prevents the instrument from pausing indefinitely if for
some reason a measurement is unable to complete.
The measurement time out is set with the command:
SYSTem:TIMeout {<time>|MINimum|MAXimum|INFinity|DEFault}
SYSTem:TIMeout? [{MINimum|MAXimum|DEFault}]
(query form)
-
time
is the specified time out in the range of
10 ms
to
2,000s
with
1 ms
resolution. The time out setting can affect reading throughput when multiple
readings are taken during the trigger cycle. Trigger delays, gate times, and gate
delays (Chapter 5) should be considered when setting an optimum time out
value.
The time out setting is stored in non-volatile memory and
is not
changed fol-
lowing a reset (
*RST
) or instrument preset (
SYSTem:PREset
or
Preset
key).
Summary of Contents for 53220A
Page 10: ...10 53220A 53230A User s Guide Contents...
Page 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Page 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Page 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Page 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Page 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Page 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...