130
53220A/53230A User’s Guide
4
53220A/53230A Input Signal Conditioning
Specifying an Absolute Threshold Level
The input threshold level can be specified as an
absolute
value. An absolute
level is set with the command:
INPut[{1|2}]:LEVel[{1|2}][:ABSolute] {<volts>|MINimum|
MAXimum|DEFault}
INPut[{1|2}]:LEVel[{1|2}][:ABSolute]? [{MINimum|MAXimum|
DEFault}]
(query form)
Absolute
threshold levels for the input ranges are:
•
5V range: ± 5.125V (2.5 mV resolution)
•
50V range: ± 51.25V (25 mV resolution
•
500V range (w/10:1 probe): ± 512.5V (250 mV resolution)
LEVel
/
LEVel1
sets the absolute threshold for all measurements except
rise/fall time and single-channel time interval. For rise/fall time measure-
ments,
LEVel
/
LEVel1
sets the
lower
reference and
LEVel2
sets the
upper
reference. For single-channel time interval,
LEVel
/
LEVel1
sets the absolute
threshold for the
start
event and
LEVel2
sets the threshold for the
stop
event.
Setting an absolute level
disables
the counter’s auto-level function.
CONFigure
and
MEASure
enable auto-leveling and set the threshold to 50%. A
reset (
*RST
) or front panel preset (
Preset
) also enables auto-leveling and sets
the threshold to 50%.
Summary of Contents for 53220A
Page 10: ...10 53220A 53230A User s Guide Contents...
Page 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Page 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Page 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Page 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Page 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Page 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...