53220A/53230A Math, Graphing, and Data Logging
The offset value (
B
) is set with the command:
CALCulate[1]:SCALe:OFFSet {<offset>|MINimum|MAXimum|
DEFault}
CALCulate[1]:SCALe:OFFSet? [{MINimum|MAXimum|DEFault}]
(query form)
offset
values can range from
-1.0E+15
to
-1.0E-15
,
0.0
, +
1.0E-15
to
+1.0E+15
The default
offset
is
0.0
and is the value set following a reset (
*RST
) or
instrument preset (
SYSTem:PRESet
).
Inverting the Reading (1/x)
Each reading as it is taken can be inverted
prior
to its use in the Mx-B equation which thus becomes
M/x-B
. Inversion is
enabled with the command:
CALCulate[1]:SCALe:INVert {OFF|ON}
CALCulate[1]:SCALe:INVert?
(query form)
ON
-
inverts the reading.
OFF
disables reading inversion.
Following a reset (
*RST
) or instrument preset (
SYSTem:PRESet
) reading
inversion is disabled.
N O T E
When
Invert x
is enabled (On) the front panel reading units (Hz, sec) are turned
off. See “Assigning Reading Units” if displayed units are required.
Summary of Contents for 53220A
Page 10: ...10 53220A 53230A User s Guide Contents...
Page 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Page 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Page 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Page 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Page 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Page 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...