104
53220A/53230A User’s Guide
3
53220A/53230A Measurements
The commands used to make phase measurements are:
MEASure:PHASe? [<channel_pair>]
CONFigure:PHASe [<channel_pair>]
FORMat:PHASe {AUTO|POSitive|CENTered}
FORMat:PHASe?
(query form)
-
channel_pair
settings are
(@1),(@2)
and
(@2),(@1)
. The
first
channel
of the pair is relative to the
second
channel.
-
AUTO
automatically selects whether the phase difference will be returned as
a positive value between 0
°
and 360
°
, or a positive or negative value between
+180
°
.
-
POSitive
returns the measured phase difference as a positive value
between 0
°
and 360
°
.
-
CENTered
returns the measured phase difference as a positive or negative
value b180
°
.
Phase Measurement Examples
// using MEASure? - measure the phase difference between
// channels 1 and 2
*RST
// reset to start from known state
FORM:PHAS CENT
// display result as pos or neg
MEAS:PHAS? (@1), (@2)
---------------------------------------------------------------------------------------------------------
// using CONFigure - measure the phase difference between
// channels 1 and 2, display result as a value between 0 and
// 360 degrees
*RST
// reset to start from known state
CONF:PHAS (@1), (@2)
FORM:PHAS POS // display result as a positive value
READ?
Summary of Contents for 53220A
Page 10: ...10 53220A 53230A User s Guide Contents...
Page 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Page 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Page 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Page 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Page 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Page 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...