
122
53220A/53230A User’s Guide
4
53220A/53230A Input Signal Conditioning
Input Impedance
The input impedance of the 53220A/53230A counter can be set to either 50
Ω
or 1 M
Ω
using the command:
INPut[{1|2}]:IMPedance {<impedance>|MINimum|MAXimum|
DEFault}
INPut[{1|2}]:IMPedance? [{MINimum|MAXimum|DEFault}]
(query form)
The 50
Ω
and
1 M
Ω
impedances allow for impedance matching (termination)
and
bridging applications respectively.
Using Probes
If using a 1:1 or 10:1 probe, the input impedance of the counter
should be set to match the compatibility (e.g. 50
Ω
, 1 M
Ω
, high Z) of the probe.
See “Setting the Probe Factor” for more information.
Following a reset (
*RST
) or front panel preset (
Preset
), the impedance is set to
1 M
Ω
.
CONFigure
and
MEASure
do not change the input impedance setting.
Impedance Example
//period measurement of expected 10 MHz signal,
//maximum resolution, use channel 1
CONF:FREQ 0.1E6,MAX,(@1)
INP:IMP 1.0E6
//set impedance to 1 Mohm
Summary of Contents for 53220A
Page 10: ...10 53220A 53230A User s Guide Contents...
Page 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Page 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Page 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Page 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Page 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Page 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...