53220A/53230A User’s Guide
247
Formats and Data Flow
7
*RST;*CLS
// reset, start from known state
SYST:TIM .001
// set a 1 ms measurement timeout
CONF:FREQ 1E6, 0.1, (@1)
// configure measurements
TRIG:COUN 5
// send 5 system triggers
SAMP:COUN 250E3
// take 250k readings/trigger
FORM:DATA REAL, 64
// set data format to binary
DATA:POIN:EVEN:THR 10E3
// set memory threshold to 10k
INIT
// initiate readings
// loop until all readings complete
For (cnt=0;cnt<1.25E6;cnt +=10000)
Do
{spoll=STAT:OPER:COND?;
// serial poll condition reg
sleep (100)
// wait 100 ms
}while !(spoll&4096) // repeat until threshold is reached
R? 10E3
// retrieve readings from memory 10k at a time
Counter File System
Measurements in volatile reading memory and counter configuration states
can be stored in files within the counter’s
internal
(non-volatile) flash mem-
ory, or on a
external
USB memory device connected to the ‘host’ port on the
front panel (Figure 7-1).
Files are created within folders on the root directory, or within sub-folders rel-
ative to the root directory. The relationship of folders and files in internal
flash memory and on the USB drive is shown in Figure 7-2.
Summary of Contents for 53220A
Page 10: ...10 53220A 53230A User s Guide Contents...
Page 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Page 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Page 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Page 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Page 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Page 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...