53220A/53230A User’s Guide
99
53220A/53230A Measurements
3
// using CONFigure - measure
positive width
at a 1.0V
// reference
*RST
// reset to start from known state
CONF:PWID 1.0V, (@1)
INP:COUP DC // set DC coupling
INP:IMP 50
// set input impedance to 50 ohm
READ?
---------------------------------------------------------
// using CONFigure - measure
negative width
at -500 mV
// reference
*RST
// reset to start from known state
CONF:NWID -500 MV, (@1)
INP:COUP DC // set DC coupling
INP:IMP 50
// set input impedance to 50 ohm
READ?
Notes
1. The reference (threshold) can be specified as a percentage of the
peak-to-peak amplitude or as an absolute level in either the
CONFigure
or
MEASure?
command.
2. Absolute references are relative to the amplitude of the signal and must take
into consideration the counter’s input coupling and impedance settings.
3. See Chapter 4 for information on configuring the input signal path which
includes coupling and impedance.