53220A/53230A User’s Guide
135
53220A/53230A Input Signal Conditioning
4
The
frequency
parameter values are:
A
frequency
setting of 10 kHz decreases the auto-level period for all frequen-
cies > 10 kHz. A
frequency
setting of 50 Hz provides auto-leveling for fre-
quencies down to 50 Hz. The
frequency
selected should be based on the
lowest
expected frequency in the measurement.
The auto-level frequency is a
system
setting which applies to all counter chan-
nels, and should be noted when making two-channel measurements.
The default setting is 50 Hz. The value is stored in non-volatile memory and
does not change when power is cycled or following a reset (
*RST
) or a front
panel preset (
Preset
).
CONFigure
and
MEASure
enable auto-level and set a
relative threshold of 50%, but do not change the auto-level frequency.
System Auto-Level Range Example
CONF:FREQ 100E6, (@1)
// measure 100 MHz signal
SYST:ALEV:FREQ 10.0E3
// set auto-level frequency
INP:IMP 50
// set impedance to 50 ohms
INP:RANG 5
// set range to 5V
INP:COUP AC
// set AC coupling
INP:LEV:REL 65
// set 65% relative threshold
Interface
Frequencies
> 10 kHz
Frequencies
50 Hz to < 10 kHz
Front Panel
Fast
Slow
SCPI
10.0E3
Maximum
50.0
MINimum
Summary of Contents for 53220A
Page 10: ...10 53220A 53230A User s Guide Contents...
Page 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Page 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Page 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Page 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Page 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Page 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...