53220A/53230A User’s Guide
95
53220A/53230A Measurements
3
MEASure:FTIMe? [{<lower_ref>|MINimum|MAXimum|DEFault}
[,{<upper_ref>|MINimum|MAXimum|DEFault}]] [,<channel>]
CONFigure:FTIMe [{<lower_ref>|MINimum|MAXimum|DEFault}
[,{<upper_ref>|MINimum|MAXimum| DEFault}]] [,<channel>]
-
lower_ref
and
upper_ref
specify the input signal reference level, either in
terms of
percent of peak-to-peak voltage
, or in
absolute voltage
. To specify
rise or fall time percentage levels
from
10%
to
90%, use a numeric value with
no
suffix or with the suffix PCT (e.g. 30 or 30 PCT).
To specify rise or fall time levels in absolute voltage within the ranges:
5.125V range: -5.125V to +5.125V
51.25V range: -51.25V to +51.25V
use a numeric value with the V or MV (millivolt) suffix: 100 MV or .1V (spaces
allowed).
If either
lower_ref
or
upper_ref
is omitted or specified in percent,
auto-leveling is enabled. If specified in absolute voltage, auto-leveling is dis-
abled
-
channel
is counter channel 1 or 2 specified as
(@1)
or
(@2)
.
Rise and Fall Time Measurement Examples
// using MEASure? - measure
rise time
using 20% and 80%
// references
*RST
// reset to start from known state
MEAS:RTIM? 20, 80, (@2)
Summary of Contents for 53220A
Page 10: ...10 53220A 53230A User s Guide Contents...
Page 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Page 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Page 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Page 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Page 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Page 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...