53220A/53230A User’s Guide
211
53220A/53230A Math, Graphing, and Data Logging
6
•
enabling limit checking
- CALCulate[1]:LIMit[:STATe] ON
•
a new measurement cycle
- INITiate:IMMediate
,
READ?
,
MEASure?
•
sending a new SCPI command or changing a current SCPI parameter
•
a reset or instrument preset
-
*RST, SYSTem:PRESet
These steps also clear
all
readings
in reading memory. (
*RST
and
SYSTem:PREset
additionally reset the lower and upper limits to 0.0.)
To turn off the ‘Limit’ annunciator and clear only the limit-sensing bits (11/12
in the Condition register)
without
clearing reading memory, the following
command is used:
CALCulate[1]:LIMit:CLEar[:IMMediate]
Histograms
The distribution for a given set of counter measurements (
except continuous
totalize and time stamp measurements
) can be represented graphically using
the 53220A/ 53230A histogram feature. The following example (Figure 6-5)
represents the basic format of a counter histogram.
Figure 6-5. 53220A/53230A Histogram Structure.
1
4
3
2
5
6
7
8
9
10
11
12
13
14
15
count
UR
LR
Bins
Rdgs < LR
Rdgs > UR
Bins = classes = points
LR = Histogram lower range boundary
UR = Histogram upper range boundary
Summary of Contents for 53220A
Page 10: ...10 53220A 53230A User s Guide Contents...
Page 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Page 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Page 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Page 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Page 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Page 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...