![Aegis OSC-9200 User Manual Download Page 78](http://html1.mh-extra.com/html/aegis/osc-9200/osc-9200_user-manual_2860410078.webp)
63
Pulse Trigger
Trigger on the positive or negative pulse with a specified width.
1.
Press
Trigger Setup
on the front panel to enter the TRIGGER function menu.
2.
Press the
Type
F key; turn the
Universal Knob
to select
Pulse
and then push the knob to confirm.
3.
Press the
Source
F key; turn the
Universal Knob
to select CH1 or CH2 as the trigger source.
4.
Press
Trigger Level
and turn the
Universal Knob
to adjust the trigger level to the desired place.
5.
Press the
Polarity
F key to select
Positive
or
Negative
pulse to trigger on. The current trigger
polarity is displayed at the upper right corner of the screen.
6.
Press the
Limit Range
F key; turn the
Universal Knob
to select the desired condition.
l
<= (less than a time value):
trigger when the input signal's positive or negative pulse
time is lower than the specified time value.
For example, for a positive pulse, if you set t (pulse real width)
≤
100ns, then the first pulse
below will not fire the trigger, but the second pulse will.
l
>= (greater than a time value):
trigger when the input signal's positive or negative pulse
time is greater than the specified time value.
For example, for a positive pulse, if you set t (pulse real width)
≥
100ns, the first pulse below
will not satisfy the trigger, but the second one will.
l
[--,--] (within a range of time value):
trigger when the input signal's positive or negative
pulse time is greater than the specified lower limit of time and lower than the specified upper
limit of time value.
For example, for a positive pulse, if you set t (pulse real width)
≥
100ns and t
≤
300ns, the
waveform will trigger.
Summary of Contents for OSC-9200
Page 1: ...AEGIS Shaft Voltage Tester Handheld Digital Oscilloscope AEGIS OSC 9200...
Page 18: ...3 Appearance and Dimensions mm Figure 1 Front View Figure 2 Side view...
Page 44: ...29 Horizontal Control Contents u Horizontal Scale u Horizontal Delay u Roll Mode u Zoom Mode...
Page 56: ...41 Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Page 60: ...45 Figure 21 With Random Noise Normal Mode Figure 22 With Random Noise Average Mode...
Page 105: ...90...
Page 135: ...120 Figure 75 Integral with Offset...
Page 139: ...124 Figure 77 Measure Pulse Width...
Page 190: ...175 DC Current Measurement AC Current Measurement...