56
Trigger Holdoff
Trigger holdoff can be used to stably trigger complex waveforms (such as pulse series). The holdoff
time is the amount of time that the oscilloscope waits before re-arming the trigger circuitry. The
oscilloscope will not trigger until the holdoff time expires.
Use the holdoff to trigger repetitive waveforms with multiple edges (or other events) between
waveform repetitions. You can also use holdoff to trigger on the first edge of a burst when you know
the minimum time between bursts.
For example, to get a stable trigger on the repetitive pulse burst below, set the holdoff time to be >200
ns but <600 ns.
The correct holdoff setting is typically slightly less than one repetition of the waveform. Set the
holdoff to this time to generate a unique trigger point for a repetitive waveform. Only edge trigger
and serial trigger have the holdoff option. The holdoff time of the oscilloscope is adjustable from
80ns to 1.5s.
1.
Press
Stop
, and then use the
Horizontal Position
button and the
Horizontal Scale
button to
find where the waveform repeats. Measure this time using cursors; then, set the holdoff.
2.
Press
Trigger Setup
to enter the TRIGGER function menu. The default trigger type is edge.
3.
Press the
Holdoff Close
F key; and then turn the
Universal Knob
to set the desired holdoff time.
Summary of Contents for OSC-9200
Page 1: ...AEGIS Shaft Voltage Tester Handheld Digital Oscilloscope AEGIS OSC 9200...
Page 18: ...3 Appearance and Dimensions mm Figure 1 Front View Figure 2 Side view...
Page 44: ...29 Horizontal Control Contents u Horizontal Scale u Horizontal Delay u Roll Mode u Zoom Mode...
Page 56: ...41 Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Page 60: ...45 Figure 21 With Random Noise Normal Mode Figure 22 With Random Noise Average Mode...
Page 105: ...90...
Page 135: ...120 Figure 75 Integral with Offset...
Page 139: ...124 Figure 77 Measure Pulse Width...
Page 190: ...175 DC Current Measurement AC Current Measurement...