Reference Position
The reference position setting determines the physical point that the oscilloscope uses during vertical
and horizontal scale changes. In some situations, it is more convenient to use a fixed position on the
display.
Press
Utility
on the front panel and then press the
Reference Pos.
F key to enter the Reference POS
menu
.
Press the
Vertical
F key to select
Fixed Offset
or
Fixed Position
.
Fixed Position
: (Default) The oscilloscope will keep the vertical offset level indicator stationary when
the vertical gain is changed.
Fixed offset
: The oscilloscope will have the vertical offset level indicator move with the actual voltage
level when the vertical gain is changed.
Press the
Horizontal
F key to select
Fixed Delay
or
Fixed Position
.
Fixed Position
: (Default) When the timebase changes, the oscilloscope will keep the horizontal offset
indicator stationary.
Fixed Delay
: When the timebase changes, the oscilloscope will have the horizontal offset indicator
move with the trigger point.
Summary of Contents for OSC-9200
Page 1: ...AEGIS Shaft Voltage Tester Handheld Digital Oscilloscope AEGIS OSC 9200...
Page 18: ...3 Appearance and Dimensions mm Figure 1 Front View Figure 2 Side view...
Page 44: ...29 Horizontal Control Contents u Horizontal Scale u Horizontal Delay u Roll Mode u Zoom Mode...
Page 56: ...41 Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Page 60: ...45 Figure 21 With Random Noise Normal Mode Figure 22 With Random Noise Average Mode...
Page 105: ...90...
Page 135: ...120 Figure 75 Integral with Offset...
Page 139: ...124 Figure 77 Measure Pulse Width...
Page 190: ...175 DC Current Measurement AC Current Measurement...