176
Recorder
The AEGIS-OSC-9200 includes a Sample Logger and Measure Logger, which record, respectively,
waveform and Measure data over time. The loggers are mainly useful for relatively low-frequency
signals. The sample logger can capture at most 25000 measurements per second, and the measure
logger can record 10 per second. The loggers will not efficiently capture high frequency signals.
Press
Recorder
to enter the RECORDER function menu.
Figure 110 Recorder Menu
Sample Logger
The Sample logger can record points from waveforms in real time at equal intervals to allow long-term
observation of low frequency signals. The recorded data can be stored in the internal storage or on an
external storage device. After stopping recording, users can replay the waveform on the oscilloscope or
export the recorded data for analysis on a PC.
Press the
Sample Logger
F key to enter the sample logger interface. At this time, most of the buttons
on the front panel will not respond, except for the menu F keys,
Meter
,
Scope
,
Hide Menu
and the
Universal Knob
. Press the
Exit
F key to exit the sample logger.
Summary of Contents for OSC-9200
Page 1: ...AEGIS Shaft Voltage Tester Handheld Digital Oscilloscope AEGIS OSC 9200...
Page 18: ...3 Appearance and Dimensions mm Figure 1 Front View Figure 2 Side view...
Page 44: ...29 Horizontal Control Contents u Horizontal Scale u Horizontal Delay u Roll Mode u Zoom Mode...
Page 56: ...41 Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Page 60: ...45 Figure 21 With Random Noise Normal Mode Figure 22 With Random Noise Average Mode...
Page 105: ...90...
Page 135: ...120 Figure 75 Integral with Offset...
Page 139: ...124 Figure 77 Measure Pulse Width...
Page 190: ...175 DC Current Measurement AC Current Measurement...