24
Adjusting the Scale
The vertical scale can be adjusted in
Coarse
or
Fine
mode.
l
Coarse
adjustment (take counterclockwise as an example): set the vertical scale in 1-2-5 step
namely 2 mV/div, 5 mV/div, 10 mV/div …10 V/div.
Fine
adjustment: further adjust the vertical scale within a relatively smaller range to improve
vertical resolution. For example: 2 V/div, 1.98V/div, 1.96V/div, 1.94 V/div …1 V/div.
l
If the amplitude of the input waveform is too large at the current scale, but would be too small if
the next scale up was used, fine adjustment can be used to improve the amplitude.
Press
CH1
on the front panel, then press the
Adjust
F key to choose the desired mode. Press
Vertical
Scale
to adjust the vertical scale (
V
to increase the scale and
mV
to reduce).
The scale information in the channel information bar at the bottom of the screen will change during
the adjustment process. The adjustable range of the vertical scale is related to the probe attenuation
ratio. By default, the probe attenuation factor is 1:1 (1X) and the adjustable range of the vertical scale
is from 2 mV/div to 100 V/div.
Summary of Contents for OSC-9200
Page 1: ...AEGIS Shaft Voltage Tester Handheld Digital Oscilloscope AEGIS OSC 9200...
Page 18: ...3 Appearance and Dimensions mm Figure 1 Front View Figure 2 Side view...
Page 44: ...29 Horizontal Control Contents u Horizontal Scale u Horizontal Delay u Roll Mode u Zoom Mode...
Page 56: ...41 Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Page 60: ...45 Figure 21 With Random Noise Normal Mode Figure 22 With Random Noise Average Mode...
Page 105: ...90...
Page 135: ...120 Figure 75 Integral with Offset...
Page 139: ...124 Figure 77 Measure Pulse Width...
Page 190: ...175 DC Current Measurement AC Current Measurement...