60
Note
: Press the Auto Setup button will set the trigger type to Edge and slope
to rising.
Edge Trigger
Edge triggering distinguishes the trigger points by seeking the specified edge (rising, falling,
alternating) and trigger level.
1.
Press
Trigger Setup
on the front panel to enter the TRIGGER system function menu.
2.
Press the
Type
F key; turn the
Universal Knob
to set select
Edge
and then push the knob to
confirm.
3.
Press the
Source
F key; turn the
Universal Knob
to select the desired trigger source.
4.
Press the
Slope
F key; turn the
Universal Knob
to select the desired trigger edge (rising, falling
or alternating), and then press down the knob to confirm. The current trigger slope is displayed
at the bottom of the screen.
5.
Press the
Trigger Level
button and turn the
Universal Knob
to adjust the trigger level to obtain
a stable trigger.
Figure 27 Edge Trigger
Holdoff, coupling, and noise rejection can be set in edge trigger mode. See the sections “Trigger
Holdoff,” "Trigger Coupling," and "Noise Rejection" for details.
Summary of Contents for OSC-9200
Page 1: ...AEGIS Shaft Voltage Tester Handheld Digital Oscilloscope AEGIS OSC 9200...
Page 18: ...3 Appearance and Dimensions mm Figure 1 Front View Figure 2 Side view...
Page 44: ...29 Horizontal Control Contents u Horizontal Scale u Horizontal Delay u Roll Mode u Zoom Mode...
Page 56: ...41 Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Page 60: ...45 Figure 21 With Random Noise Normal Mode Figure 22 With Random Noise Average Mode...
Page 105: ...90...
Page 135: ...120 Figure 75 Integral with Offset...
Page 139: ...124 Figure 77 Measure Pulse Width...
Page 190: ...175 DC Current Measurement AC Current Measurement...