R3765/67G Series Network Analyzer Operation Manual
5.7 High-speed Measurement Using the Program Sweep Function
5-41
•
Next, enlarge and measure the specified bandwidth about this filter by using program
sweep.
Here, divide the pass-band into the following three segments to enlarge and measure.
Each sweep segment (SEG.) is independent so that different sweeping point number,
power level value and IF bandwidth value can be setup.
Program sweep function allows to sweep up to all sweeping point 1201 at a time and
connect up to 30 kinds of these segment sweep settings (SEG.NUMBER 0 to 29).
Figure 5-35 Setting of Program Sweeping
SEG.
START
STOP
POINT
0
800MHz
840MHz
50
1
865MHz
900MHz
50
2
920MHz
960MHz
50