R3765/67G Series Network Analyzer Operation Manual
7.5 Calibration
7-46
{P1 ~ P4 THRU.STD}
: Enters the parameters of the through standard to be
connected between Port 1 to Port 4.
Calls the offset entry menu. (See step
.)
* This is available when OPT11 (Built-in 3-port test set)
or OPT14 (Built-in 4-port test set) is installed.
{P2 ~ P4 THRU.STD}
: Enters the parameters of the through standard to be
connected between Port 2 to Port 4.
Calls the offset entry menu. (See step
.)
* This is available when OPT11 (Built-in 3-port test set)
or OPT14 (Built-in 4-port test set) is installed.
{P3 ~ P4 THRU.STD}
: Enters the parameters of the through standard to be
connected between Port 3 to Port 4.
Calls the offset entry menu. (See step
.)
* This is available when OPT11 (Built-in 3-port test set)
or OPT14 (Built-in 4-port test set) is installed.
{SAVE TO USER DEFINE}: Saves the parameters entered.
calls the saving menu. (See step
.)
Open standard entry menu
{OPEN C0}
: Enters the open capacitance, C
0
.
{OPEN C1}
: Enters the open capacitance, C
1
.
{OPEN C2}
: Enters the open capacitance, C
2
.
{OPEN C3}
: Enters the open capacitance, C
3
.
{OPEN OFFSET}
: Enters the electric length of the open standard.
Calls the offset entry menu. (See step
.)
{SHORT OFFSET}
: Enters the electric length of the short standard.
Calls the offset entry menu. (See step
.)
Offset entry menu
{OFFSET DELAY}
: Enters the electric length of the open/short/through
standard.
{OFFSET LOSS}
: Enters the loss of the through standard.
{OFFSET Z0}
: Enters the impedance of the through standard.
Saving menu
{YES}
: Saves the parameters of the calibration kit.
{NO}
: Does not save the parameters of the calibration kit.
7
7
7
8
6
7
7
7
8