R3765/67G Series Network Analyzer Operation Manual
A.3 Initial Setting
A-7
*
2 : Start/Stop level
*
3: Selection item of input port
*
4: Measurement format
Type
Power on or preset
*RST
Start
Stop
Start
Stop
AG
-3dBm 0dBm
-3dBm 17dBm
AG+OPT10
-8dBm 0dBm
-8dBm 12dBm
BG
-13dBm
0dBm
-13dBm
7dBm
BG+OPT10
-18dBm 0dBm
-18dBm 2dBm
BG+12
-11dBm
0dBm
-11dBm
9dBm
BG+OPT10+OPT12
-16dBm
0dBm
-16dBm
4dBm
CG, CG+OPT11/14
-10dBm
0dBm
-10dBm
10dBm
CG+OPT12/13
-16dBm 0dBm
-16dBm 4dBm
CG+OPT10,
CG+OPT10+OPT11/14
-15dBm 0dBm
-15dBm 5dBm
CG+OPT10+OPT12/13
-21dBm -1dBm
-21dBm -1dBm
Type
Channel
CH1
CH2
CH3
CH4
AG
A/R
B/R
A/R
B/R
BG
REFLECTION
TRANSMISSION
REFLECTION
TRANSMISSION
CG
S11
S21
S11
S21
Type
Channel
CH1
CH2
CH3
CH4
AG
LOG MAG
LOG MAG
LOG MAG
LOG MAG
BG
LOG MAG
LOG MAG
POLAR
LOG MAG
CG
LOG MAG
LOG MAG
POLAR
LOG MAG