R3765/67G Series Network Analyzer Operation Manual
7.5 Calibration
7-63
(31-C) Connect the load standard between TEST PORT2 and TEST PORT3, then press
{P2-P3 ISOLATION}.
A message "Wait for Sweep ." is displayed and the calibration data is acquired.
When the message disappears, the calibration data acquisition is completed.
(Note)
(31-D)Connect the load standard between TEST PORT1 and TEST PORT4, then press
{P1-P4 ISOLATION}.
A message "Wait for Sweep ." is displayed and the calibration data is acquired.
When the message disappears, the calibration data acquisition is completed.
(Note)
(31-E) Connect the load standard between TEST PORT2 and TEST PORT4, then press
{P2-P4 ISOLATION}.
A message "Wait for Sweep ." is displayed and the calibration data is acquired.
When the message disappears, the calibration data acquisition is completed.
(Note)
(31-F) Connect the load standard between TEST PORT3 and TEST PORT4, then press
{P3-P4 ISOLATION}.
A message "Wait for Sweep ." is displayed and the calibration data is acquired.
When the message disappears, the calibration data acquisition is completed.
(Note)
(31-G)Press {DONE ISOLATION} to execute the isolation calibration.
When the isolation calibration has been completed, the 4-port full calibration menu
is displayed.
Note:
It is possible to retrieve calibration data for each calibration standard if
{DONE ISOLATION} has not been pressed yet.
Press {DONE 4-PORT} to execute the 4-port full calibration.
32