R3765/67G Series Network Analyzer Operation Manual
10.9 Dynamic Level Accuracy
10-52
10.9.2 Dynamic Level Accuracy for the R3765BG/67BG
This section describes the detail test procedure of the dynamic level accuracy. This section
applies only to the R3765BG/67BG.
10.9.2.1 Specifications
Reference level :
-19 dBm
(-24 dBm for OPT 10, -28 dBm for OPT 10 + OPT 12, -23 dBm for OPT 12)
0 dB to -10 dB :
±
0.20 dB (300kHz to 3.8GHz)
0 dB to -10 dB :
±
0.40 dB (3.8GHz to 8GHz)
-10 dB to -50 dB :
±
0.05 dB
-50 dB to -60 dB :
±
0.10 dB
10.9.2.2 Instruments Required
•
Step attenuator (Range: 0 to 90 dB, Accuracy within 0.02dB)
Recommended model: HP8496B
•
RF cable (SMA(m)/SMA(m) 50
Ω
) quantity=2
Recommended model: A01253-060
•
Adopter (N(m)/SMA(f)) quantity=2
Recommended model: HRM-554S
•
3-dB attenuator (SMA(f)/SMA(m)) quantity=2
Recommended model: AT-103
•
50
Ω
to 75
Ω
conversion adapter (75
Ω
N(m)/50
Ω
N(f))
×
2: 11852B (For OPT 12)
10.9.2.3 Procedure
(1)
Connect two 3-dB attenuators and step attenuator using two RF cables between TEST
PORT 1 and TEST PORT 2 shown in Figure 10-31.
Figure 10-31 Connections for TEST PORT 1 Measurement
for Dynamic Level Accuracy
* Connect the 50
Ω
to 75
Ω
conversion adapter to
TESTPORT 1 and 2 for
OPT 12.
Note that the 3 dB fixed
attenuator is not required.