R3765/67G Series Network Analyzer Operation Manual
7.9 Sweep
7-95
7.9.3
Editing Segment of Program Sweep
The Setting and the Explanation
Press the [MENU] to call the signal source menu. (Refer to section A.4.)
Press the {SWEEP TYPE[ ]} to call the sweep type menu.
Press the {EDIT PROG SWEEP} to call the program sweep segment editing menu.
Program sweep segment editing menu
•
Program sweep segment editing menu (1 of 2)
{SEGMENT: NUMBER}
: Specifies the segment number in the range of 0 to 29.
{START}
: Sets the start frequency of the specified segment.
{STOP}
: Sets the stop frequency of the specified segment.
{POINT}
: Sets the point number of the specified segment.
{CLEAR SEG}
: Clears the specified segment.
{CLEAR ALL SEG}
: Clears all segments.
•
Program sweep segment editing menu (2 of 2)
{SEGMENT: POWER}
: Sets the output level of the specified segment.
{IF RBW}
: Sets the receiver section resolution bandwidth of the
specified segment.
{SETTLING TIME}
: Sets the settling time of the specified segment.
CAUTION!
1. If the same segment number is edited on the segment editing menu of the program
sweep, then the user frequency segment is also changed. (Segment holds the pro-
gram sweep in common.)
2. The total number of points of each segment by program sweep cannot exceed 1201
points. (The maximum number of measurement points is 1201 points.)
1
2
3
4