R3765/67G Series Network Analyzer Operation Manual
7.5 Calibration
7-37
(3)
Normalize & isolation calibration
The Setting and the Explanation
Set the instrument to the transmission measurement mode.
Press [CAL] to call the calibration menu (1/2).
Press {CAL MENU} to call the calibration selection menu.
Press {2 PORT CAL MENU} to call the 2-port calibration selection menu.
Press {NORMALIZE & ISO’N} to call the normalization and isolation calibration menu.
Connect the through standard between the test ports and press {THRU}.
The message “Wait for Sweep .” is displayed and the calibration data is obtained. The
calibration is completed when the message disappears. (Note)
Connect the load standard to each test port and press {ISOLATION}.
The message “Wait for Sweep .” is displayed and the calibration data is obtained. The
calibration is completed when the message disappears. (Note)
Press {DONE NORM & ISO} to complete the calibration of the normalize & isolation
calibration.
Connect DUT to perform the measurement.
Note:
Do not move the R3765/67G series, the cable, the connector, the standard, and
the others when the message “Wait for Sweep .” is displayed.
When the set condition is changed during the message is displayed, the
message “Calibration aborted .” is displayed and the current calibration data
cannot be acquired.