79
Create the corresponding DUT tasks as described above, and fulfill the necessary information. See Table 34.
Table 33 Loop readout task setting
Subject
Valid value
Description
Name
Numbers, letters, symbols
Task name
S/N
Numbers, letters, symbols
Task S/N
Model
Numbers, letters, symbols
Model of DUT
Range
(4~20)mA,(0~10)mA,(0~20)mA,(1~5)V,(0~5)V,(0~10)V, custom
Input range of DUT
Accuracy
0.025%,0.05%,0.1%,0.16%,0.25%,0.4%,1%,1.6%,2.5%,4%,custom
Accuracy of DUT
Owner
Numbers, letters, symbols
Owner of DUT
Location
Numbers, letters, symbols
Location of DUT
Note
Numbers, letters, symbols
Additional information
3. Task setting
Refer to the task setting in paragraph 6.1.1.1.
4. Start the task
For part of the content, refer to Start the task in paragraph 6.1.1.4.
During the task performing, the calibrator will output the corresponding electrical signal according to the set
point and the DUT input type (current/voltage) set by the user. For the measured value at each set point, users
need to manually input them.
5. Finish the task
Refer to Finish the task in paragraph 6.1.1.5.