(3) Repeat step 1 and 2 until the whole task is finished.
5. Finish the task
After the task is finished, the task report will automatically come up. In the task report page, you can view the detailed
data of each set point. Click the save button
at the bottom right of the interface to fulfill in the information required,
and save the task data of the result.
Table 22 Save task data setting
Subject
Valid value
Description
Operator
Numbers, letters, symbols
Information for operator
Date and time
Valid date and time
Date and time of the task performing
Temperature
Number(
℃
as unit)
Ambient temperature when task performing
Humidity
Number(%RH as unit)
Ambient humidity when task performing
Save as
As found/ As left/ Both
Result type
6.1.2 Pressure transmitter
1. Connect to DUT
Connect the external pressure module to the Lemo type connection A or B.
Connect the DUT output with calibrator’s electrical module according to DUT output signal type. (current: figure
7/voltage: figure 6/HART: figure 10 or 11)
2. Create the task
Create the corresponding DUT tasks as described above, and fulfill the necessary information.
Summary of Contents for 226
Page 1: ...226 227 Multifunction Process Calibrator...
Page 2: ......
Page 3: ...ADT226 227 Multifunction Process Calibrator User Manual Version 2306V01 Additel Corporation...
Page 4: ......
Page 6: ......
Page 31: ...19 1 4 Basic Structure Figure 1 Basic Structure...
Page 33: ...21 Figure 2 Power adaptor...
Page 64: ...52 External module information including external module A and external module B...
Page 107: ...95...