![Additel 226 User Manual Download Page 88](http://html2.mh-extra.com/html/additel/226/226_user-manual_3047281088.webp)
76
Table 31 Flow volume totalizers task setting
Subject
Valid value
Description
Name
Numbers, letters, symbols
Task name
S/N
Numbers, letters, symbols
Task S/N
Model
Numbers, letters, symbols
Model of DUT
Frequency
Numbers
Frequency of DUT input signal
Amplitude
Numbers
Amplitude of DUT input signal
Flow unit
Option
Flow unit of DUT, variety of flow units available
Decimals
Option 0~3
The last decimal digit of the DUT’s display
Total flow
Numbers
Expected total flow when the task finish
K-Param
Numbers
K coefficient of DUT (The flow rate represented by each signal pulse)
Tolerance
Numbers
Tolerance of DUT
Owner
Numbers, letters, symbols
Owner of DUT
Location
Numbers, letters, symbols
Location of DUT
Note
Numbers, letters, symbols
Additional information
3. Task setting
Refer to the task setting in paragraph 6.1.1.1.
4. Start the task
For part of the content, refer to Start the task in paragraph 6.1.1.4.