3.5.9 Automatic step output
The three electrical signal output gears of voltage, current, and frequency of the calibrator and the two simulate
output gears all support automatic step output mode. In the automatic step output mode, the calibrator outputs signals in
sequence according to the output points set by the user. The automatic step function provides flexible setting items to
control the step output behavior and create different types of output points. As shown in Table 18.
Table 17 Automatic step output settings
Subject
Valid Value
Description
Stroke type
One way/ round
Select stroke type
Cycle times
(
0~100
)
Automatic step cycle times, if set 0, it means infinite loop
Cycle interval
(
0~999999
)
Interval time between cycles
Dwell time
(
0~999999
)
Dwell time between step points
Step mode
Step point/ percentage/ engineering
unit/ custom
Select step mode
Number of step points
(
2~16
)
Number of step points, available for Step point mode
Step size
Percentage: 7~100
Engineering unit: dynamic range
Step size, available for percentage/ engineering unit
Step point list
Step point
Add or delete set points, available for custom step mode
Scale
Numbers
Lower and upper point of the automatic step scale
Summary of Contents for 226
Page 1: ...226 227 Multifunction Process Calibrator...
Page 2: ......
Page 3: ...ADT226 227 Multifunction Process Calibrator User Manual Version 2306V01 Additel Corporation...
Page 4: ......
Page 6: ......
Page 31: ...19 1 4 Basic Structure Figure 1 Basic Structure...
Page 33: ...21 Figure 2 Power adaptor...
Page 64: ...52 External module information including external module A and external module B...
Page 107: ...95...