U
SER
M
ANUAL
750000992/16
9-11
MTS
e
Command relative to the OTDR module
Format des commandes
Description des fonctions
Note of up to 41 characters associated with an event. For each wavelength, a
maximum of 16 notes are possible.
Command relative to the OTDR module
Acquisition commands
Format des commandes
Description des fonctions
LASer
Sets the laser source to the appropriate wavelength supported by
the optics subsystem of the unit (depending on the module
installed).
PULSe
Sets the pulse width to a value supported by the optics subsystem
of the unit (from 3 ns to 20 µs depending on the modules).
RANGe
Sets the distance range to a value supported by the optics
subsystem of the unit (depending on the modules).
MAXTime
Sets the acquisition period used for averaged acquisitions (from 5
s. to 10 mn).
Note: a real time acquisition will stop after 10 minutes.
RESolution
Set the resolution to a value supported by the optics subsystem of
the unit (from 8 cm to 320 m depending on the module):
ONOTes
[1:100], [string of 41characters]
ACQuisition
RANGe
MAXTime
PROGram
[R1, R2,
…
R8] The range depends on the module
(see page 11-13)
[5 : 600]
[MANual, AUTO]
LASer
[L850, L1300, L1310, L1550, L1625]
PULSe
[P3NS, P5NS, P10Ns, P20Ns, P30Ns, P50Ns, P100ns,
P200ns, P300ns, P500ns, P1US, P3US, P10Us,
RESolution
[AUTO, MANU].[0.0:320.0]
BEEP
[ON, OFF]