11-4
U
SER
M
ANUAL
750000992/16
Technical specifications
MTS
e
OTDR Measurement characteristics
Distance measurement
•
dual cursors
•
distance displayed with respect to the calibration of the fiber’s group index
•
group index adjustable from 1.30000 to 1.70000 in 0.00001 step
•
display resolution: 1 cm (0.39 inch) max
•
cursor resolution: 1cm (0.39 inch) max
•
sampling spacing: from 4 cm (1.57 inch), with up to 128 000 acquisition data
points
•
accuracy:
±
1 m
±
10
-5
x distance
±
sampling resolution
(excluding group index uncertainties)
Attenuation measurement
•
dual cursors
•
display resolution: 0.001 dB
•
cursor resolution: 0.01 dB
•
accuracy:
±
0.05 dB / dB
±
0.05 dB
Reflectance measurement
•
display resolution: 0.01 dB
•
accuracy:
±
4 dB
Automatic measurement
•
Automatic measurement of all signal characteristics. Slope measurement by
linear regression or 2 points measurements.
•
Thresholds for fault display:
0 to 5.99 dB in 0.01 dB steps for event threshold
-11 to -99 dB in 1 dB steps for reflectance
0 to 1.99 dB/km in 0.001 dB/km steps for slope
•
Display of fiber segment slope and attenuation
•
Display of position and loss
•
Display of reflectance
•
Display of ORL
Manual measurement
•
measurement of slope between the cursors
•
measurement of the loss between two fiber sections
•
measurement of the reflectance of a reflective event
•
measurement of the Optical Return Loss between the two cursors.
•
measurement of a splice either by 2 point or 5 point method
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com