U
SER
M
ANUAL
750000992/16
3-11
MTS
e
After switching on
Press the YES key.
"
Programming flash memory; in progress
": the MTS
e
then programs its
internal memory and starts up the new MTS
e
software.
"
Completed
" appears then the screen is blank: the software upgrade is
completed. The MTS
e
is ready for use when the LED
ON
is flashing.
6.
Press the
ON
key to switch it on.
Instrument Set up
The instrument can be equipped with up to two modules and with several test
functions. Before making measurements, it is necessary to select the function or
functions required in the Instrument Set-up menu.
If a module has been changed, after power up, the unit will directly go to this
Instrument Set-up menu.
The Talk Set option can be selected even if an OTDR function is selected. OTDR,
OSA, VFL and Loss Test Set (Power Meter & Laser Source) functions are
exclusive: the selection of one of them will cancel the previous selection.
To configure the instrument
Press the
SET-UP
button : the configuration menu is displayed with the last type
of configuration defined: Test, System or Instrument.
If the configuration menu displayed is not the one required, press the
<Instrument Setup>
key.
The functions of the instrument are graphically displayed : OTDR, OSA, Source,
Power Meter or Visual Fault Locator installed according to the modules present
in the MTS
e
.
To validate a function:
•
Use the directional arrows to move to the function icon.
•
Press the
Select
button to validate the choice. The icon will then be
highlighted (in red with color screen) and the connection to the relevant
connector will be displayed.
If none of the function is selected
•
The MTS
e
is in the OTDR mode if an OTDR module is present, but no
acquisition can be made. You can however analyze, store and recall traces.
•
The MTS
e
is in the WDM mode if an OSA module is present but no OTDR
module.