Using the OTS module
MTS
e
OTS TEST SETUP window (including light source and power meter)
In these windows, the selected parameter is high-lighted. Select this parameter
using the
↓
and
↑
directional arrows.
The available choices are displayed on the screen; these depend on the chosen
function. Make the required choice with the
←
and
→
directional arrows.
Selecting Light source parameters
You can choose the following Light source parameters, from the
Light Source
menu of the
OTS Test Setup
window:
Wavelength
The required source:
1310 nm or 1550 nm for LASER source,
850 or 1300 nm for LED sources.
Mode
Choose the transmission mode:
CW
: Continuous signal.
270 Hz, 330 Hz, 1 KHz
and
2 KHz
: the source transmits a
modulated signal at the frequency selected.
Emul. OLS
: to communicate with an Acterna OLS or OFI
instrument.
Scanning
The source then switches every 10 seconds between the two
wavelengths and provides a signal indicating the
wavelength of the transmitted signal. When the Scanning