MTS
e
Table of results
3.
Use the
<
Event
>
key. The measurements given are the true distance from the
beginning of the link, and the attenuation of the link plus the attenuation of
the connection.
This function may also be used to determine the loss budget for a partial section
of the fiber under test.
The
<
Event
>
key is directly linked to the Table of Results which is described
below.
Table of results
Two types of tables of results are available with the MTS
e
:
•
A single line table below the full trace display giving the type and
characteristics of the event nearest the current cursor on the trace.
•
A eight line table listing the events around the current cursor. The event line
nearest the cursor is highlighted. The highlighted line is updated if you move
the cursor or select the next/previous event.
Notes
In order to display all the events, verify that the thresholds are set to
ALL
in the
TEST SET UP
menu.
In order to display the results on the trace, verify that the
Results on
trace
option is activated in the
TEST SET UP
menu.
To see this table, in the Results window, press the <Table> key. To scroll the table
if it has more than 7 lines, either move the cursor, or press the <Event> key then
the
↑
or
↓
keys.
Note
Each event is identified on the trace by an order number.