5. Setting up the device
11
5.4. Extended parameterisation
(Professional operation level)
Menu group level
Menu level
5.4.1. Signal input parameters
Setting the end value of the measuring range,
END
:
Default:
10000
Set the end value from the smallest to the highest digit with
[
▲
] [
▼
]
and confirm each digit with
[P]
. A minus sign can only be parameterized on the highest value digit. After the last digit, the
display switches back to the menu level. If a sensor calibration was selected as input option,
you can only select between
noca
and
cal.
With
noca
, only the previously set display value is
taken over, and with
cal
, the device takes over both the display value and the analogue input
value.
Setting up the start/offset value of the measuring range,
offs
:
Default:
0
Enter the start/offset value from the smallest to the highest digit with
[
▲
] [
▼
]
and confirm each
digit with
[P]
. After the last digit the display switches back to the menu level. If a sensor
calibration was selected as input option, you can only select between
noca
and
cal.
With
noca
,
only the previously set display value is taken over, and with
cal
, the device takes over both the
display value and the analogue input value.
Selection of the input signal,
tYPE:
Default:
s.60ou
Available as measuring input options are 0-600/0-300 VAC or 0-5/0-1 AAC signals as works
calibration (without application of the sensor signal) and
S.600v, S.300v, se.5a
and
se.1a
as sensor
calibration (with the sensor applied). Select with
[
▲
] [
▼
]
and confirm the selection with
[P]
.
Parameterisation level
Menu level