What You See on the Display
What You Do
Comments
7-2
Questions? Refer to Page 10-10 for Technical Support Instructions.
5.
Select Glucose or
β-
Ketone Linearity
Panel.
If
β-
Ketone testing is disabled, this screen
will not appear.
6.
Scan or manually enter the CVC kit lot
number via the keypad, then press
Enter.
LIN stands for Linearity.
7.
If the
New Panel
screen appears you may
either:
Press
1
–
ReEnter Kit Lot.
Press
2
–
Replace Panel.
If the CVC kit lot number that was entered
is different from the previous one, the
monitor will prompt you to do one of two
actions:
Either re-enter the existing lot number or
replace the existing lot number with the new
lot number.
8.
Select the number of the level of the next
test to be run. If you press
6
for a
New
Panel
, the monitor will prompt you to
confirm that you wish to replace the
existing panel.
There can be up to 4 replicates per level.
The number in the parentheses indicates
the number of replicates already run for that
level. When a level is full, the number 4 will
appear in parentheses, indicating that all
four replicates have been completed for this
level. For example
1 - Level 1(4)
.
Chapter 7: Linearity Test