Issue
Solution
i
‑
STAT Alinity only displays the last 3 numbers or
letters of the Operator ID or No Operator ID. The
entire Operator ID should display.
Check the Operator ID Presentation feature in the
Operator ID Entry feature on the profile assigned
to the instrument. Follow CWi instructions in this
manual to change the display.
i
‑
STAT Alinity only displays the operator’s first
name and last initial. The entire name should
display.
Check the Operator Name Presentation in the
Operator ID Entry feature on the profile assigned
to the instrument. If the check box is marked for
Partially Display Operator Name, uncheck it to
display entire name. Follow CWi instructions in
this manual.
i
‑
STAT Alinity will not upload operator list
Check and confirm that the Search for Operator ID
on Operator List feature in the Operator Settings
Category is checked on the profile assigned to the
instrument. If not checked, follow CWi instructions
in this manual to change the profile and enable
the instrument to load the operator list from the
data management system. Check the Connectivity
Settings Category to confirm the IP address is
programmed for operator list.
i
‑
STAT Alinity will not upload cartridge lot list
Check to confirm Search Cartridge List for
Cartridge Lot Number is enabled in the Quality
Settings Category on the profile assigned to the
instrument. If not enabled, follow CWi instructions
in this manual to enable Search Cartridge List for
Cartridge Lot Number to load the cartridge list
from the data management system. Check the
Connectivity Settings Category to confirm the IP
address is programmed for cartridge list.
Patient armband barcode will not scan
Check Barcode settings in the Patient Settings ID
Entry feature on the profile assigned to the
instrument. Follow CWi instructions in this manual
to change the barcode type.
Patient list will not upload onto the i
‑
STAT Alinity
Check and confirm that the Search Patient List for
Patient ID feature is enabled in the Patient
Settings Category on the profile assigned to the
instrument. If it is not, follow the CWi instructions
in this manual to change the profile and enable
the instrument to load the Patient List from the
data management system. Check the Connectivity
Settings Category to confirm the IP address is
programmed for Patient List.
252
i-STAT Alinity — System Operations Manual
Art: 746304-01 Rev. H
Rev. Date: 29-Oct-2019
Summary of Contents for i-STAT Alinity
Page 1: ......
Page 12: ...xii i STAT Alinity System Operations Manual Art 746982 01 Rev E Rev Date 02 Nov 2022 ...
Page 24: ...xxiv i STAT Alinity System Operations Manual Art 746983 01 Rev H Rev Date 11 Nov 2020 ...
Page 39: ......
Page 44: ...6 i STAT Alinity System Operations Manual Art 745526 01 Rev H Rev Date 04 Mar 2021 ...
Page 48: ...10 i STAT Alinity System Operations Manual Art 745527 01 Rev N Rev Date 13 Oct 2022 ...
Page 84: ...46 i STAT Alinity System Operations Manual Art 745531 01 Rev G Rev Date 04 Mar 2021 ...
Page 88: ...50 i STAT Alinity System Operations Manual Art 745529 01 Rev G Rev Date 04 Mar 2021 ...
Page 96: ...58 i STAT Alinity System Operations Manual Art 745530 01 Rev D Rev Date 16 May 2022 ...
Page 100: ...62 i STAT Alinity System Operations Manual Art 745578 01 Rev C Rev Date 04 Mar 2021 ...
Page 108: ...70 i STAT Alinity System Operations Manual Art 745525 01 Rev E Rev Date 04 Mar 2021 ...
Page 110: ...72 i STAT Alinity System Operations Manual Art 745532 01 Rev D Rev Date 04 Mar 2021 ...
Page 134: ...96 i STAT Alinity System Operations Manual Art 745524 01 Rev I Rev Date 02 Nov 2022 ...
Page 147: ......
Page 164: ...126 i STAT Alinity System Operations Manual Art 747079 01 Rev F Rev Date 28 Apr 2020 ...
Page 167: ......
Page 176: ...136 i STAT Alinity System Operations Manual Art 746298 01 Rev M Rev Date 13 Oct 2022 ...
Page 250: ...210 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 290: ...250 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 296: ...256 i STAT Alinity System Operations Manual Art 746304 01 Rev H Rev Date 29 Oct 2019 ...