Quality Check Failure Resolutions
Resolutions
A
Navigate to the Home Screen, then touch More Options. Touch Instrument Status and assess
the instrument's temperature. Move the instrument to an appropriate environment.
B
Perform an Electronic Simulator test. If the test results in a PASS, the instrument is ready for use
otherwise contact system administrator for further instruction.
C
Repeat Electronic Simulator testing. If the test results in a PASS, the instrument is ready for use
otherwise contact system administrator for further instruction.
D
Contact the system administrator for further instruction.
E
Do not collect the sample for this cartridge in a device that contains anticoagulant. Draw new
sample. Repeat testing with a freshly filled cartridge. Carefully observe the help provided
throughout the testing pathway. If the same quality check failure displays, contact the system
administrator for further instruction.
F
Prepare a new bottle of material per the manufacturer's instruction. Repeat testing with a
freshly filled cartridge. Carefully observe the help provided throughout the testing pathway. If
the same quality check failure displays, contact the system administrator for further instruction.
G
Repeat testing with a freshly filled cartridge. Carefully observe the help provided throughout the
testing pathway. If the same quality check failure displays, contact the system administrator for
further instruction.
H
When filling a cartridge, use care to advance blood to the level indicated by the 'fill to' arrow.
Repeat testing with a freshly filled cartridge. Carefully observe the help provided throughout the
testing pathway. If the same quality check failure displays, contact the system administrator for
further instruction.
I
Instrument did not reset correctly. Perform an electronic simulator test. If the test results in a
PASS, the instrument is ready for use otherwise contact system administrator for further
instruction.
J
The simulator was not fully inserted. Repeat testing. Make sure that the cover retaining ring
does not interfere with Electronic Simulator insertion. Ensure the Simulator is fully inserted. The
Simulator is fully inserted when the click is heard. If the same quality check failure displays,
contact the system administrator for further instruction.
K
Always scan the barcode found on the pouch that contained the cartridge in use. Scanning any
other barcode can cause this error. Repeat testing with a freshly filled cartridge. Carefully
observe the help provided throughout the testing pathway. If the same quality check failure
displays, contact the system administrator for further instruction.
L
A cartridge was detected when an Electronic Simulator was expected. Repeat testing ensuring to
insert an Electronic Simulator. Make sure that the cover retaining ring does not interfere with
Simulator insertion. Ensure the Simulator is fully inserted. The Simulator is fully inserted when
the click is heard. If the same quality check failure displays, contact the system administrator for
further instruction.
M
Reserved for future use.
38
i-STAT Alinity — System Operations Manual
Art: 745527-01 Rev. N
Rev. Date: 13-Oct-2022
Summary of Contents for i-STAT Alinity
Page 1: ......
Page 12: ...xii i STAT Alinity System Operations Manual Art 746982 01 Rev E Rev Date 02 Nov 2022 ...
Page 24: ...xxiv i STAT Alinity System Operations Manual Art 746983 01 Rev H Rev Date 11 Nov 2020 ...
Page 39: ......
Page 44: ...6 i STAT Alinity System Operations Manual Art 745526 01 Rev H Rev Date 04 Mar 2021 ...
Page 48: ...10 i STAT Alinity System Operations Manual Art 745527 01 Rev N Rev Date 13 Oct 2022 ...
Page 84: ...46 i STAT Alinity System Operations Manual Art 745531 01 Rev G Rev Date 04 Mar 2021 ...
Page 88: ...50 i STAT Alinity System Operations Manual Art 745529 01 Rev G Rev Date 04 Mar 2021 ...
Page 96: ...58 i STAT Alinity System Operations Manual Art 745530 01 Rev D Rev Date 16 May 2022 ...
Page 100: ...62 i STAT Alinity System Operations Manual Art 745578 01 Rev C Rev Date 04 Mar 2021 ...
Page 108: ...70 i STAT Alinity System Operations Manual Art 745525 01 Rev E Rev Date 04 Mar 2021 ...
Page 110: ...72 i STAT Alinity System Operations Manual Art 745532 01 Rev D Rev Date 04 Mar 2021 ...
Page 134: ...96 i STAT Alinity System Operations Manual Art 745524 01 Rev I Rev Date 02 Nov 2022 ...
Page 147: ......
Page 164: ...126 i STAT Alinity System Operations Manual Art 747079 01 Rev F Rev Date 28 Apr 2020 ...
Page 167: ......
Page 176: ...136 i STAT Alinity System Operations Manual Art 746298 01 Rev M Rev Date 13 Oct 2022 ...
Page 250: ...210 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 290: ...250 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 296: ...256 i STAT Alinity System Operations Manual Art 746304 01 Rev H Rev Date 29 Oct 2019 ...