Not all cartridges are available in all regions. Check with your local representative for
availability in specific markets.
Contact pads
The contact pads conduct the signals generated by the sensors to the instrument. In order for these to
function properly, care must be exercised not to contaminate the contact pads during cartridge handling.
Sensors
The sensors are electrodes microfabricated on silicon chips located within the cartridge. Electrodes have
chemically sensitive coatings such as ion-selective membranes and enzyme layers. Each sensor is
connected to a contact pad by a signal line. The sensors respond to the calibrant solution and the sample
by producing measurable signals related to analyte concentration.
Sensor Channel
The sensor channel directs the sample from the sample chamber to the sensors. An extension of this
channel becomes a waste chamber to receive the calibrant solution as it is displaced by the sample.
Air Chamber
An air chamber is located in blood gas/electrolyte/chemistry/hematocrit cartridges between the sample
chamber and sensor channel. This creates an air segment between the calibrant solution, if applicable,
and the sample to prevent the two from mixing. The size of the air segment is monitored by the
instrument.
Calibrant Pack
During the first part of the testing cycle, the calibrant solution is automatically released from the foil
pack and is positioned over the sensors. The foil pack may be pierced with improper cartridge handling.
If fluid is released prematurely, a Quality Check Failure may occur.
Fill To
The blue arrow on the white label cartridge and the clear arrow on the blue cover cartridge is intended
to help the operator fill the cartridge correctly. Samples that fall above or below the indicated level may
result in a Quality Check Failure.
Sample Chamber
The sample chamber includes the sample well and the channel leading from the well up to the fill mark.
When filled, the sample chamber contains sufficient sample for testing. Sample volume and placement
are monitored by the instrument.
Bladder
The bladder is connected to the sample well. The instrument presses on the bladder to displace calibrant
solution from the sensors, to move the sample from the sample chamber to the sensors, or to mix
sample and reagents.
Sample Well
The area on the cartridge where the sample is introduced into the sample chamber.
90
i-STAT Alinity — System Operations Manual
Art: 745524-01 Rev. I
Rev. Date: 02-Nov-2022
Summary of Contents for i-STAT Alinity
Page 1: ......
Page 12: ...xii i STAT Alinity System Operations Manual Art 746982 01 Rev E Rev Date 02 Nov 2022 ...
Page 24: ...xxiv i STAT Alinity System Operations Manual Art 746983 01 Rev H Rev Date 11 Nov 2020 ...
Page 39: ......
Page 44: ...6 i STAT Alinity System Operations Manual Art 745526 01 Rev H Rev Date 04 Mar 2021 ...
Page 48: ...10 i STAT Alinity System Operations Manual Art 745527 01 Rev N Rev Date 13 Oct 2022 ...
Page 84: ...46 i STAT Alinity System Operations Manual Art 745531 01 Rev G Rev Date 04 Mar 2021 ...
Page 88: ...50 i STAT Alinity System Operations Manual Art 745529 01 Rev G Rev Date 04 Mar 2021 ...
Page 96: ...58 i STAT Alinity System Operations Manual Art 745530 01 Rev D Rev Date 16 May 2022 ...
Page 100: ...62 i STAT Alinity System Operations Manual Art 745578 01 Rev C Rev Date 04 Mar 2021 ...
Page 108: ...70 i STAT Alinity System Operations Manual Art 745525 01 Rev E Rev Date 04 Mar 2021 ...
Page 110: ...72 i STAT Alinity System Operations Manual Art 745532 01 Rev D Rev Date 04 Mar 2021 ...
Page 134: ...96 i STAT Alinity System Operations Manual Art 745524 01 Rev I Rev Date 02 Nov 2022 ...
Page 147: ......
Page 164: ...126 i STAT Alinity System Operations Manual Art 747079 01 Rev F Rev Date 28 Apr 2020 ...
Page 167: ......
Page 176: ...136 i STAT Alinity System Operations Manual Art 746298 01 Rev M Rev Date 13 Oct 2022 ...
Page 250: ...210 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 290: ...250 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 296: ...256 i STAT Alinity System Operations Manual Art 746304 01 Rev H Rev Date 29 Oct 2019 ...