1.1 - Complete i
‑
STAT Alinity System Overview
The i
‑
STAT Alinity instrument is an analytical, in vitro diagnostic device. The instrument requires i
‑
STAT
single-use cartridges containing electrodes and sensors to perform quantitative diagnostic testing on
whole blood or plasma. Together, the instrument and cartridge allow the user to perform clinical testing
and related administrative tasks.
The i
‑
STAT Alinity design enables the instrument to be taken to the patient’s bedside (point of care), a
convenient location near the point of care, or clinical laboratory setting.
After the insertion of a filled test cartridge, the instrument carefully monitors and controls the testing
process. The only user intervention is in the form of data entry, performed via the touchscreen or by
barcode capture. Throughout the cycle, the instrument performs a series of quality checks. These checks
are designed to monitor the status of the instrument and the quality of the cartridge. An i
‑
STAT Alinity
instrument, a cartridge with the required test, and two or three drops of blood will allow the caregiver to
view quantitative test results within minutes.
For the purposes of this system operations manual, the i
‑
STAT Alinity system components and associated
accessories will be discussed. Additional information related to the entire system can be found in the
i
‑
STAT Alinity system documentation listed below.
Note: Note Regarding System Reliability: The i-STAT System automatically runs a
comprehensive set of quality checks of analyzer and cartridge performance each time a
sample is tested. This internal quality system will suppress results if the analyzer or
cartridge does not meet certain internal specifications (see Quality Control section in
System Operations Manual for detailed information). To minimize the probabilty of
delivering a result with medically significant error, the internal specifications are very
stringent. It is typical for the system to suppress a very small percentage of results in
normal operation given the stringency of these specifications. If however, the analyzer or
cartridges have been compromised, results may be persistently suppressed, and one or
the other must be replaced to restore normal operating conditions. Where unavailability
of results while awaiting replacement of analyzers or cartridges is unacceptable, APOC
recommends maintaining both a back up i-STAT System instrument and cartridges from
an alternate lot number.
i
‑
STAT Alinity Documentation:
•
i
‑
STAT Alinity System Operations Manual, including:
○
i
‑
STAT Alinity Reference
○
i-STAT Cartridge Instructions for Use
○
AlinIQ NCi - Network Connectivity for i
‑
STAT Alinity
○
AlinIQ CWi - Customization Workspace for i
‑
STAT Alinity
•
i
‑
STAT Alinity Quick Reference Guide
•
i
‑
STAT Alinity Getting Started Guides:
○
i
‑
STAT Alinity Base Station
○
i
‑
STAT Alinity Rechargeable Battery
○
i
‑
STAT Alinity Electronic Simulator
○
i
‑
STAT Alinity Printer
3
i-STAT Alinity — System Operations Manual
Art: 745526-01 Rev. H
Rev. Date: 04-Mar-2021
Summary of Contents for i-STAT Alinity
Page 1: ......
Page 12: ...xii i STAT Alinity System Operations Manual Art 746982 01 Rev E Rev Date 02 Nov 2022 ...
Page 24: ...xxiv i STAT Alinity System Operations Manual Art 746983 01 Rev H Rev Date 11 Nov 2020 ...
Page 39: ......
Page 44: ...6 i STAT Alinity System Operations Manual Art 745526 01 Rev H Rev Date 04 Mar 2021 ...
Page 48: ...10 i STAT Alinity System Operations Manual Art 745527 01 Rev N Rev Date 13 Oct 2022 ...
Page 84: ...46 i STAT Alinity System Operations Manual Art 745531 01 Rev G Rev Date 04 Mar 2021 ...
Page 88: ...50 i STAT Alinity System Operations Manual Art 745529 01 Rev G Rev Date 04 Mar 2021 ...
Page 96: ...58 i STAT Alinity System Operations Manual Art 745530 01 Rev D Rev Date 16 May 2022 ...
Page 100: ...62 i STAT Alinity System Operations Manual Art 745578 01 Rev C Rev Date 04 Mar 2021 ...
Page 108: ...70 i STAT Alinity System Operations Manual Art 745525 01 Rev E Rev Date 04 Mar 2021 ...
Page 110: ...72 i STAT Alinity System Operations Manual Art 745532 01 Rev D Rev Date 04 Mar 2021 ...
Page 134: ...96 i STAT Alinity System Operations Manual Art 745524 01 Rev I Rev Date 02 Nov 2022 ...
Page 147: ......
Page 164: ...126 i STAT Alinity System Operations Manual Art 747079 01 Rev F Rev Date 28 Apr 2020 ...
Page 167: ......
Page 176: ...136 i STAT Alinity System Operations Manual Art 746298 01 Rev M Rev Date 13 Oct 2022 ...
Page 250: ...210 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 290: ...250 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 296: ...256 i STAT Alinity System Operations Manual Art 746304 01 Rev H Rev Date 29 Oct 2019 ...