interfering substance in the sample
•
aged sample containing products of metabolism
•
cartridge results that are not reportable based on a failed quality check while performing a test
If stars display, take corrective action:
•
A star out can be caused if a cartridge is handled improperly by the user. Follow facility policy for
handling starred out results.
•
If no facility policy is in place, check the supply of cartridges in use with a control solution. If the
control is in range, collect a new sample and test immediately. If starred out results display again,
there may be an interfering substance in the sample. Refer to the Cartridge Instructions for Use for a
list of interfering substances with the starred out analyte(s). Use an alternate method for testing.
•
If the control is out-of-range or if stars are displayed again, there may be a problem with the cartridge
lot. Use another lot if available. If another lot is not available, test patient sample using another
method. Contact technical support with regard to lot number in question.
Suppressed Results - Out of range
Under some conditions results may be preceded by the symbols for greater than (>) or less than (<). This
can occur when results are outside the system’s measurement ranges for the analyte . Facility Policy
should be in place to advise end users on additional course of action if necessary. Below is an example of
a result that is greater than the highest limit of the reportable range (180 mmol/L). The result is shown
as a greater than (>) symbol and the upper limit for the analyte:
Suppressed Results - Null Set
The < > flag indicates that the results for this test were dependent on the result of a test flagged as
greater than (>) or less than (<) reportable range of the test. A null result is shown in the example below.
Null results don't include a numerical value:
Interpreting Results with customization features applied
The following section provide examples for various analytes, when the analyzer is customized with Align
IQ CWI. Not all cartridges are available in all regions. Ability to run a cartridge is controlled by the
instrument and by region.
44
i-STAT Alinity — System Operations Manual
Art: 745531-01 Rev. G
Rev. Date: 04-Mar-2021
Summary of Contents for i-STAT Alinity
Page 1: ......
Page 12: ...xii i STAT Alinity System Operations Manual Art 746982 01 Rev E Rev Date 02 Nov 2022 ...
Page 24: ...xxiv i STAT Alinity System Operations Manual Art 746983 01 Rev H Rev Date 11 Nov 2020 ...
Page 39: ......
Page 44: ...6 i STAT Alinity System Operations Manual Art 745526 01 Rev H Rev Date 04 Mar 2021 ...
Page 48: ...10 i STAT Alinity System Operations Manual Art 745527 01 Rev N Rev Date 13 Oct 2022 ...
Page 84: ...46 i STAT Alinity System Operations Manual Art 745531 01 Rev G Rev Date 04 Mar 2021 ...
Page 88: ...50 i STAT Alinity System Operations Manual Art 745529 01 Rev G Rev Date 04 Mar 2021 ...
Page 96: ...58 i STAT Alinity System Operations Manual Art 745530 01 Rev D Rev Date 16 May 2022 ...
Page 100: ...62 i STAT Alinity System Operations Manual Art 745578 01 Rev C Rev Date 04 Mar 2021 ...
Page 108: ...70 i STAT Alinity System Operations Manual Art 745525 01 Rev E Rev Date 04 Mar 2021 ...
Page 110: ...72 i STAT Alinity System Operations Manual Art 745532 01 Rev D Rev Date 04 Mar 2021 ...
Page 134: ...96 i STAT Alinity System Operations Manual Art 745524 01 Rev I Rev Date 02 Nov 2022 ...
Page 147: ......
Page 164: ...126 i STAT Alinity System Operations Manual Art 747079 01 Rev F Rev Date 28 Apr 2020 ...
Page 167: ......
Page 176: ...136 i STAT Alinity System Operations Manual Art 746298 01 Rev M Rev Date 13 Oct 2022 ...
Page 250: ...210 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 290: ...250 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 296: ...256 i STAT Alinity System Operations Manual Art 746304 01 Rev H Rev Date 29 Oct 2019 ...